> Troubleshoot - RENASYS GO
Troubleshooting - RENASYS GO
Having trouble with your RENASYS GO device? Follow these troubleshooting guidelines below for help.
The RENASYS GO device is very noisy. Check the following
-
Check for any air leaks at the dressing, port, connector or canister
- If the RENASYS GO device is having to pump more air than normal it will pump more frequently and the sound level will be higher
- Disconnect the dressing from the canister and cap the canister tubing. The device should fall virtually silent. If the device does not fall silent, change the O-ring on the pump inlet. For instructions click here.
-
Check that the vacuum set point is correct and has not been set higher than prescribed
- The higher the vacuum set point the harder the pump must work and the greater the sound level.
> Technical Support UK: 0800 915 5394 Ireland: 1800 3036 22 <