Having trouble with your RENASYS GO device? Follow these troubleshooting guidelines below for help.
Extremely Low Battery
Switch off the RENASYS GO device and charge the battery via the external power supply for 3 hours.
Battery indicator light should slowly flash green indicating charging is underway and will be solid green when charging is complete.
Do not power the RENASYAS GO device on until approximately 10 minutes of charging has occurred, i.e. until the battery has recovered from the deep discharge.